I. Key Features
Lightweight, aesthetically pleasing, with a wide current range.
Section II: Key Technical Specifications
1Environmental Conditions
Please provide the Chinese content to be translated. Degree:5—40℃ Moisture Level:80(25℃)
Elevation: 1500m Power Frequency:50Hz
2Rated Capacity:2.5-80VA (O-3.2Ω)
3Measurement Range:5%-120%
4Secondary Current: 5A,1A
5Power Factor:0.8,1.0
6Accuracy: ±3
7Test Voltage:2kV

User's Guide
1Connect this device to the test current transformer.K2With the calibratorTxBetween.
2Adjust the load to the actual secondary load of the test current transformer.
Note:Secondary loads of some current transformers are represented by resistances.,Unit: Ω.The load of a current transformer burden box is generally expressed in terms of capacity.VA.Their conversion relationship is::
Capacity(VA)=Resistance Value(Ω)*Secondary Current(A)* Secondary Current(A)
(1Ensure proper grounding of this equipment.
Caution: To ensure the accuracy of the secondary load of the transformer, it is necessary to connect the secondary terminals of the transformer to the two connection lines of the load box, with a total resistance value of0.06If the secondary reconnection verification meter is used, the load caused by the secondary of the transformer should also be taken into account to ensure accuracy.