PFH02 is compact in size and structure, with an all metal pump head and an anodized aluminum casing. It is suitable for high-precision filling applications with small flow rates, with a filling volume as low as 30uL (accuracy better than 1%). The phase compensation design reduces pulsation and improves filling accuracy
Flow range: - ≤ 1178ml/min
Larger number of channels: 1
Applicable locations:
Support desktop small desktop packaging and can be matched with automated/semi-automatic filling machinery
Product Features
Compact size and structure
All metal pump head, shell made of anodized aluminum material
Small flow high-precision filling application, with a filling volume as low as 30uL (accuracy better than 1%)
Phase compensation design reduces pulsation and improves filling accuracy
Features:
Compatible Drivers: Suitable for dPOFLEX PFU, dPOFLEX PFS, and most existing driver products (with nine hole mounting brackets)
Applicable hose types: The product supports silicone tubing (hose assembly) and Biosicon medical silicone tubing (hose assembly)
Applicable hose wall thickness: 2.4mm
Applicable hose component specifications: PFH02-06G、PFH02-08G、PFH02-12G、 PFH02-16G、 PFH02-24G、PFH02-32G 、PFH02-48G
Suitable filling needles: 0.6mm, 1.0mm, 1.6mm, 3.2mm, 4.5mm inner diameter filling needles
Filling volume range per second: 0.2ml-16ml
Filling accuracy: Within the filling volume range of 0.2ml-16ml, the filling accuracy error is<=± 1%, which can reach ± 0.5%
High speed: 450rpm
Repeatability accuracy: CV<0.5%
Noise: 50dB (tested at a distance of 1m)
Dimensions: 83mm × 83mm × 76mm
Weight: 0.508kg
Working environment: Working temperature range: 5-40 ° C, working relative humidity: relatively high 85% RH, non condensing
Transportation and storage environment: Temperature range for transportation and storage: -40-70 ° C, relative humidity for transportation and storage: relatively high 85% RH, non condensing
Technical Indicators:


Other parameters:
Pump head model | Roller material | speed (rpm) | Inside the hose (mm) | Large reference flow rate (mL/min) | Small filling volume (accuracy ≤± 1%) | Large second filling volume | Recommended inner diameter of filling needle (mm) |
PFH02 | stainless steel | ≤450 | 0.6 | 27 | 0.2mL | 0.42mL | 0.6 |
0.8 | 66 | 0.4mL | 1.03mL | 0.6 | |||
1.2 | 114 | 0.5mL | 1.78mL | 1 | |||
1.6 | 147 | 0.8mL | 2.29mL | 1 or 1.6 | |||
2.4 | 371 | 1.16mL | 5.77mL | 1.6 | |||
3.2 | 566 | 2.15mL | 8.80mL | 1.6 or 3.2 | |||
4.8 | 1178 | 3mL | 18.28mL | 3.2 or 4.5 |