I. Oxygen Content Analyzer Overview
1. Features wide power supply and character-based human-machine interface, power failure protection settings, password permission settings, input signal fault indication (displays "br" when input signal is disconnected), and configurable input/output, for easy user operation and maintenance.
2. Equipped with isolated RS232 or RS485 communication interfaces, compatible for communication with computers, PLCs, and DCS systems.
II. Technical Specifications
1. Usage Conditions: Environmental Temperature: 0~50℃
Relative Humidity: 90% RH
2. Power Voltage: 90-265VAC, 24±10%VDC
3. Output Characteristics: Relay Rating: 0.5A/240V AC, resistive.
Voltage transmitter output load: >100kW
Current type transmitter output load impedance: <600Ω
4. Power Consumption: <5W
Weight: <0.5kg
Section 3: Applicability Range
This gauge is suitable for measuring oxygen content and exceeding-alarm in large and medium-sized hyperbaric chambers in hospitals (note: according to the requirements of the Chinese Medical Association Hyperbaric Oxygen Society, oxygen content in hyperbaric chambers should be controlled below 25% where conditions permit, generally below 30%); it is used for continuous monitoring of oxygen content in gas during the synthesis ammonia process in fertilizer factories, and can be matched with electrostatic deasphalting equipment. It can promptly cut off the static high voltage when the oxygen content in the chemical process exceeds the set value, preventing accidents. The instrument is also applicable to industrial and mining enterprises, ship compartments, tunnels, coal mines, chemical equipment containers, water purification plants, gas, internal combustion engine exhaust, thermal electric equipment, smelting, boilers, kilns, environmental protection monitoring, oxygen generators, quality inspection, electrolytic hydrogen equipment, submarines, artificial satellites, civil defense, space travel. It is used for oxygen content measurement in grain, tobacco leaves, fruits, vegetables, and medical department ventilators, anesthesia machines, and various other situations. It can also be used for oxygen quantity control in crop and animal respiration experiments.
IV. Technical Specifications of Sensors
1. Measurement Range: 0-100%
2. Working Life: 6 years in the air
Output: 13 to 16.5 mV
Resolution: 0.1%
5. Temperature Range: 0℃ to 40℃
6. Pressure Range: 750 to 1250 hPa
7. Response Time: (T 90): < 12S
8. Humidity Range: 0-100% RH (High humidity resistance)
9. Zero Output: (100% N2, 20℃) < 0.2 mV
10. Large Zero Drift: (20℃ to 40℃): 1%
11. Long-term Drift: <1% / per month
12. Recommended Load Value: >10KΩ
13. Linearity Output: 2%
14. Repeatability: 1%
15. Interface: 3-pin socket with temperature compensation
Section 5: Sensor Outline Dimensions Diagram