Performance Features
Constant Voltage Output: The adjustment rate of the output voltage is within ±1%, preventing changes in output voltage due to unstable input power voltage and load variations, ensuring accurate measurement results.
Features rapid discharge capability; the tester can discharge the test object and the circuit within 0.2 seconds after DC testing, ensuring the safety of the operator.
Safety electric wall and protection system, which can safeguard the safety of operators and ensure that the tested items are not damaged.
Output AC voltage frequency range is selectable between 50Hz and 60Hz.
Test results can be automatically saved within the testing device for easy verification by the user.
Also, the test results can be connected to an RS232 interface and imported into an EXCEL spreadsheet for data statistics and analysis.
List display function available; upon completion of tests with more than 2 steps, the tester automatically displays the results of each step.
Standard PLC interface, RS232C interface, standard supervisory software.
The programmable high-voltage unit features an oscilloscope interface, enabling the monitoring of arc discharge and flashover phenomena of the object being tested.