Logan KeXing KB-5 Portable Liquid Particle Size Analyzer

Overview:

Offline Detection:

   Sample directly from the sample bottle for testing; connect to the pressure bottle sampling device (pressure unit) for testing.

Online Inspection:

   Installable on various hydraulic transmission, lubrication, oil filter, cleaning, and testing benches, providing online detection of system oil contamination (cleanliness).

Performance Features:

l Utilizing the principle of photoresist (light shielding) method, it boasts advantages such as fast detection speed, strong anti-interference capability, high accuracy, and good repeatability.

l High-precision sensors ensure high resolution power and accuracy.

l Precise measurement sampling system, ensuring consistent sampling speed and precise volume.Please provide the Chinese content to be translated.Control.

l 5.6Inch color LCD display, touch screen operation.

l Built-inNAS1638、ISO4406-1999、ISO4406-2017、ISO4406-1987、GB/T 14039-2002、GJB420B-2006、GJB420B-2015、GJB420A-1996、AS4059EDifferential,AS4059EAccumulatedAS4059FDifferentialAS4059FAccumulatedROCT17216-1971、ROCT17216-2001、JTS-005、SAE749D-1963、NAS1638/ISO4406Particle contamination grade standards available, and can be internally configured to meet customer requirements.

l Built-in"Customized" standard, with intervals0.1μmCan be set to any particle size.990Particle size, for easy particle size analysis.

l Built-in calibration feature, adjustableISO4402、GB/T18854-2002(ISO11171:1999)、ROCTCalibrate to meet standards.

l Automatic calibration feature, with a convenient and quick calibration process.

l Store up to three calibration curves simultaneouslyACFTDCalibration curvesISOMTDCalibration CurveGOSTCalibration curves are available and can be easily switched, minimizing conversion errors.

l Built-in printer, capable of directly printing out test reports.Enter numbers, letters, and Chinese characters when printing.

l Data Storage Function: The testing data can be stored either within the instrument's memory or via the instrument'sUSBThe interface will store the detected data in the external plug.UPlated.

l The instrument allows for continuous automatic on-line detection, providing real-time particle counts and grades, and allows for adjustable detection interval settings.

l Optional water activity sensor for detecting water activity in oilawValue and oil temperature.

l Equipped with flushing function, the flushing volume can be arbitrarily adjusted.

l High-strength injection-molded shell, compact structure, lightweight, easy to carry.

l High-precision aluminum anodized panel, sleek and aesthetically pleasing, durable for long-lasting use.

l StandardRS232Communication, and provide communication protocols, allowing external connection to computers or higher-level machines to control the instrument and transmit, store, and process detection data.

 

 

 

Technical Specifications:

l Light    Semiconductor Laser

l Test Range:1μm~600μm(Subject to the selected sensor)

l Particle size range:1~100μm( ISO4402)or4~70μm(c) (GB/T18854,ISO11171)

l Spirit Sensitivity Degree:1μm(ISO4402)or4μm(c)(GB/T18854,ISO11171)

l Inspection Channel:32Pieces, with intervals0.1μmAny particle size can be set

l Testing Method: Offline or Online

l Sample Volume:10ml/Next

l Relative error of sampling volume: better than±3%

l Counting Volume:10ml

l Inspection Time:30S/Next 

l Testing Speed:5~60mL/minDefault20mL/min)

l Washing Speed:5~60mL/minDefault20mL/min)

l Flush Volume: Available0ml~90mlIntermediate settings1ml)

l Overlap Error Limit:12000~40000Particle/mL(Subject to the selected sensor)

l Particle count repeatability:RSD<2%

l Relative Error of Particle Count±10%

l Results Storage: Within or Inside the InstrumentUPlatter

l Offline viscosity detection:≤100cStOptional pressure bottle sampler with viscosity up to400cSt)

l Online Pressure Testing0.1~0.6MpaOptional pressure reducing device available40MPa)

l Online Inspection Interval: Variable1Instantly10Hour59Please provide the Chinese content to be translated.59Instant setup

l Test Sample Temperature:0℃~80℃

l Operating Temperature:-20℃~60℃

l Storage Temperature:-30℃~80℃

l Supply    Electricity100V~265VACOr24VDCOr external lithium-ion battery

l Dimensions:415×310×170mm

l Net Weight:7kg

 

 

 

 

 

 

 


Disclaimer:Info provided by user, user liable for authenticity, accuracy & legality. Zhongshang114 assumes no liability.

Tip:Confirm supplier qualification & quality before purchase to avoid risks.

Unit Price Negotiable
Inquiry None
Delivery TianjinPre-sale, after payment15dayswithin
Brand Logan Kexing
Light Source Semiconductor laser
Testing Scope 1μm to 600μm
Particle size range 1-100 μm
Expiry Long Valid
Update 2023-04-11 15:15
Gallery

Tianjin Logan KeXing Technology Co., Ltd.Published bySupply Logan Kexing KB-5 Portable Liquid Particle Size AnalyzerGallery Lib