Equipment Brief Introduction
LYL-17LBST Series Double-Door Muffle Furnaces incorporate advanced technology and are self-developed and produced energy-saving, environmentally friendly, and new type electric furnaces. They feature an advanced and reasonable structure, and are aesthetically pleasing. The outer shell is made of high-quality cold-rolled steel plate, precision-machined on CNC machines, and finished with a luxurious and beautiful two-color imported epoxy powder electrostatic spraying process, which is resistant to high temperatures and corrosion. The heating elements are made of high-quality resistance wire/silicon carbide rod/silicon molybdenum rod, and the double-layer furnace shell with air insulation layer keeps the surface temperature of the furnace shell low. This furnace boasts even temperature distribution, low surface temperature of the furnace body, adjustable heating rate, energy-saving, and other advantages. The furnace chamber is constructed from ceramic fiber material using CNC technology, offering strong resistance to thermal shock, excellent resistance to rapid cooling and heating, good corrosion resistance, no collapse, no crystallization, no slag, no pollution, and a long service life. The control system utilizes microcomputer artificial intelligence adjustment technology, featuring...PIDAdjustable, fully automatic control, self-tuning function, multi-step programing, capable of programming various heating and temperature-holding programs, with high temperature control accuracy; integrated module with silicon-controlled rectifier control, phase-shifting triggering, and protection system: featuring independent over-temperature protection, over-pressure, over-flow, leakage, short-circuit, and other protective measures. High level of automation, with all indicators reaching advanced levels.
Primary Application
This equipment is widely used in universities, research institutions, industrial and mining enterprises, and is an ideal device for experiments and production in fields such as powder, electronics, metallurgy, pharmaceuticals, ceramics, sulfur, new glass materials, chemicals, metal sintering, and metal heat treatment.
Corporate Qualifications & Honors
Quality Management System CertificationGB/T19001:2016/ISO 9001:2015; European UnionCECertification; "Luoyang City Model Enterprise for Civilized and Trustworthy Business Operation"“China Electric Furnace Industry***”;“Quality, Service, Credibility - AAA Corporation”Our products have passed inspections by *** certification bodies, and third-party certification reports can be provided upon request.
Main Functions & Features
1. Energy-efficient and pollution-free (high-purity composite alumina/porcelain crystalline fiber for furnace material, lightweight, low heat storage, energy-saving, operating at half power on average)
2. Easy replacement and energy-saving (with high-quality resistance wire/silicon carbide rod/silicon molybdenum rod heating elements, long service life and easy replacement)
3. Integrated design for flexibility in use, less prone to tangling, and space-saving operation.
4. High temperature control accuracy (high-quality instruments, ±1℃ temperature control accuracy) with third-party inspection certificate available for the instrument.
5. Uniform heating temperature (reasonable layout, balanced thermal field, no blind spots in the furnace cavity)
6. Microcomputer program-controlled temperature regulation, adjustable heating rate, automatic operation without the need for human supervision
7. Over-temperature, over-pressure, over-flow, leakage, broken coupling, broken rod automatic protection and alarm
8. Side-hinged door for easy access
9. Wide Application Range: Universities, research institutions, industrial and mining enterprises, etc., for powder, military industry, electronics, metallurgy, pharmaceuticals, ceramics, glass, machinery, new material development, special materials, refractory materials, building materials, chemicals, metal sintering, and metal heat treatment, etc., for testing and production.
10. Double-shell design with air convection insulation layer
11. Electrostatic powder coating on the shell, with a stylish and generous two-color design
12. Excellent reputation, numerous clients
13. Great value for money, comprehensive after-sales service, 1-year warranty, lifetime on-site repair
14. Dual furnace chamber, individually controllable; one furnace can function as two.
Main Technical Parameters
Item | Indicator Value |
Rated Voltage | 220V/380V |
Design Temperature | 800℃,1000℃,1200℃,1400℃,1600℃,1700℃,1800℃ |
Rising Rate | ≤0.1-20℃/min (adjustable) |
Heating Element | High-Quality Imported Resistance Wire / High-Quality U-Type Silicon Carbon Rod / High-Quality U-Type Silicon Molybdenum Rod |
Temperature Control Accuracy | ±1℃ |
Temperature Control Method | Microcomputer control, PID regulation, intelligent microcomputer control, multi-step programmable temperature control curve, no need for supervision (fully automatic heating, insulation, power off when door is opened) |
Thermometric element | K(0-1300℃)/S(0-1600℃)/B(0-1820℃) |
Furnace chamber material | Dual-chamber design, independently controllable, one furnace can be used as two, ceramic polycrystalline fiber /Alumina fiber/Alumina Fiber/High-purity aluminum oxide composite fiber/Imported Fiber |
Furnace door structure | Front door side-hinged |
Cooling Method | Double-wall shell with air insulation layer |
Surface Temperature Rise | ≤30℃ |
Random gift | 2 heating rods, 2 rod holders, 1 crucible tongs, 1 pair of heat-resistant gloves, 1 furnace bottom plate, 1 user manual, 1 certificate of conformity |
Selection Guide
Model | Furnace chamber dimensions (depth x width x height in mm) | Furnace Volume (L) | Power Voltage (V) | Power (KW) | Temperature Control Accuracy |
LYL-17LBST-1 | 200*150*150 | 4.5 | 220 | 4 | ±1℃ |
LYL-17LBST-2 | 300*200*120 | 7.2 | 220 | 6 | ±1℃ |
LYL-17LBST-3 | 300*200*150 | 9 | 220 | 8 | ±1℃ |
LYL-17LBST-4 | 300*200*200 | 12 | 220/380 | 10 | ±1℃ |
LYL-17LBST-5 | 400*250*160 | 16 | 380 | 12 | ±1℃ |
LYL-17LBST-6 | 400*300*300 | 36 | 380 | 14 | ±1℃ |
LYL-17LBST-7 | 500*500*500 | 125 | 380 | 25 | ±1℃ |
LYL-17LBST-8 | 700*500*500 | 175 | 380 | 30 | ±1℃ |
LYL-17LBST-9 | 800*500*500 | 200 | 380 | 40 | ±1℃ |
LYL-17LBST-10 | 800*800*800 | 512 | 380 | 90 | ±1℃ |
LYL-17LBST-11 | 1200*800*800 | 768 | 380 | 100 | ±1℃ |
LYL-17LBST-12 | 1000*1000*1000 | 1000 | 380 | 120 | ±1℃ |
Temperature Category | 800°F-1800°F available for selection | ||||
Customized Non-Standard | According to customer requirements, customize furnaces in various sizes. | ||||
Standard Accessories
Name | Image | Quantity | Description | |
High-Temperature Gloves | | 1Double | Comfortable to wear and keeps hands warm; cut-resistant, abrasion-resistant, and heat-resistant (can withstand 350°C, contact heat for 60 seconds, 100°C contact heat for 180 seconds) durable. | |
Heating Element | Silicon carbide rod | | 2Branch | Silicon carbide components, with high usage temperatures, are hard and brittle. After molding, they are used, such as silicon carbide rods and tubes, which can reach high heating temperatures under air-dry conditions.1500100-120°C, for vacuum conditions; 1000-1200°C under nitrogen atmosphere; long-term use at 1350°C |
Silicon Molybdenum Rod | | 2Please provide the Chinese content to be translated. | Silicon molybdenum rods are heated to high temperatures in an oxidizing atmosphere, forming a dense quartz glass film on the surface to prevent further oxidation. Therefore, silicon molybdenum rods possess high-temperature oxidation resistance. In an oxidizing atmosphere, the temperature can reach1800°C, can be used as heating elements for industrial high-temperature furnaces in sectors such as electronics, ceramics, magnetic materials, glass, metallurgy, and refractory materials. | |
Tongs for crucibles | | 1Please provide the Chinese content that you would like translated into American English. | 1. A common chemical instrument, typically used for gripping crucibles. Generally made of stainless steel or hard, non-combustible, and non-oxidizable materials. 2. Use clean tongs. 3. When gripping a hot crucible with tongs, preheat the tip of the tongs to prevent the crucible from cracking due to localized cooling. After use, place the tongs with the tip facing up on the table or a asbestos pad. 4. After the experiment, clean the crucible tongs and place them in the laboratory equipment cabinet for drying. 5. Use the bent portion for holding crucibles; for other uses, employ the pointed tip. 6. Crucible tongs are not necessarily used in conjunction with crucibles; they can also be used separately. | |
Furnace bottom plate | | 1Piece | Ceramic furnace bottom plates, heat and abrasion resistant, with load-bearing capabilities, protect the furnace base from wear, extending the equipment's lifespan. | |
List of Optional Accessoriesnal Accessories List)
Individual control cabinet | Touchscreen Gauge | Remote Assistance Module |
Observation window | Exhaust ports and chimneys | Portable Heater Fan |
Crucibles, molds | Door block | Bottom plate |
Blower | Thermocouple | Centralized Control Industrial Computer |
Casters | Multi-point temperature measurement | Exhaust Emission Purification Equipment |
Special Note
Quality Assurance | One-year warranty on the entire unit, excluding related consumables (excluding pads and heating elements)Lifetime On-Site Maintenance |
Factory Environment | 1. Indoor use 2. Ambient temperature within ±5-45°C range. 3. Relative ambient temperature does not exceed 85%. 4. No conductive dust, explosive gases, or gases that can severely damage metals and cause insulation corrosion are present around the stove. |
Safety Protection | 1. When using an electric furnace for the first time or after a long period of disuse, it must be seasoned to prevent the furnace chamber from cracking. 2. Keep clean regularly and check the electrical furnace wiring for good contact periodically. 3. When using the electric furnace, the temperature must not exceed the rated temperature to prevent damage to the heating elements. It is prohibited to pour various flammable liquids or molten metals into the furnace. 4. During the process, the temperature should generally not exceed 300°C, and the heating rate should be moderate. This is because when heating begins, the furnace is cold and requires a significant amount of heat absorption. |
After-sales service | 1. We promise to address any technical inquiries from our clients within 1 hour. 2. We promise to issue the test results for equipment returned for repair within 1 working day, complete the repair within 3 working days, and handle special materials as needed. 3. Within one week after returning the equipment to the customer, conduct a follow-up visit with the user and provide free technical consultation. |
Friendly Reminder | Do not operate the equipment at rated temperature for extended periods to ensure its normal service life. 2. Carefully count and verify the quantity of goods and accessories upon receipt to avoid any inconvenience caused by shortages or omissions. 3. To extend the lifespan of the instrument, we recommend not exceeding a heating or cooling rate of 10℃/min. 4. The cooling rate should not exceed 5°C/min. 5. This furnace does not have a vacuum-sealed structure, thus it is not suitable for introducing flammable or explosive gases. 6. After a period of use, the furnace chamber may develop minor cracks, which is a normal occurrence and will not affect its operation. It can be repaired using an alumina coating. |
Common Faults and Troubleshooting
Fault Phenomenon | Fault Cause | Exclusion Method |
Turn on the control power switch, no power indicator displayed | Control circuit, fuse blown, power off | Inspect for short circuits, eliminate if present, and then replace the fuse. |
Equipment operates with voltage but no current. | Heating element fracture | Replace heating element |
Instrument PV display shows the upper temperature limit of the corresponding type of thermocouple. | Thermocouple short circuit | Replace Thermocouple |
Power is normal, but the furnace cannot heat up. | Control main line damage | Notify the maintenance personnel for inspection. |