Equipment Brief Introduction
The LYL-18MFL series muffle furnaces incorporate advanced technology, featuring self-developed and produced energy-saving, environmentally friendly, and new type electric furnaces with an advanced and reasonable structure, and are aesthetically pleasing. The exterior shell is made of cold-rolled steel plates, processed by CNC machine tools, and finished with a luxurious and beautiful two-color imported epoxy powder electrostatic spraying process, which is resistant to high temperatures and corrosion. It uses resistance wire/silicon carbide rod/silicon molybdenum rod as the heating element, a double-layer furnace shell air insulation layer to maintain a low surface temperature of the furnace shell. This furnace boasts an even thermal field, low surface temperature of the furnace body, adjustable heating rate, energy-saving advantages, and more. The furnace chamber is constructed from ceramic fiber material using CNC technology, with strong resistance to thermal shock, excellent resistance to rapid cooling and heating, good corrosion resistance, no collapse, no crystallization, no slagging, no pollution, and a long service life. The control system employs microcomputer artificial intelligence adjustment technology.PIDAdjustable, fully automatic control, self-tuning function, multi-programming capabilities, with programmable heating and temperature-holding cycles; high temperature control accuracy. Integrated module with thyristor control and phase-shifting triggering, with protection system featuring independent over-temperature, over-pressure, over-flow, leakage, short-circuit, and other protections. High degree of automation, with all indicators reaching advanced levels.
Primary Application
This equipment is widely used in universities, research institutions, industrial and mining enterprises, and is an ideal device for experiments and production in powder, electronics, metallurgy, ceramics, sulfur, new glass materials, chemicals, metal sintering, and metal heat treatment.
Corporate Qualifications & Honors
Quality Management System CertificationGB/T19001:2016/ISO 9001:2015; European UnionCECertification; "Luoyang City Model Enterprise for Civilization and Trustworthy Business Practice"“China Electric Furnace Industry Brands”;“Quality, Service, Reputation - AAA Corporation”Our products pass inspections by certification bodies, and third-party certification can be provided upon request.
Main Functions & Features
1. Energy-efficient and environmentally friendly (high-purity composite alumina/porcelain crystalline fiber furnace material, lightweight, low heat retention, energy-saving; typically operates at half power)
2. Easy replacement with energy and time savings (heating elements are made of resistance wire/silicon carbide rod/silicon molybdenum rod, with long service life and easy replacement)
3. Integrated design for flexibility in use, less prone to snagging, and space-saving.
4. High temperature control accuracy (high-quality instruments, ±1℃ temperature control accuracy) with third-party inspection certificate available for the instrument.
5. Excellent uniformity of furnace temperature (合理布局,温场均衡,炉腔无死角) - Good uniformity of furnace temperature (rational layout, balanced temperature field, no dead corners in the furnace cavity)
6. Microcomputer programmable temperature control with adjustable heating rate; automatic operation, no need for human supervision
7. Over-temperature, over-pressure, over-flow, leakage, unbalanced coupling, broken rod automatic protection and alarm
8. Side-opening furnace door, easy to open
9. Wide Application Range: Universities, research institutions, industrial and mining enterprises, etc., for powder, military, electronics, metallurgy, ceramics, glass, machinery, new material development, special materials, refractory materials, construction materials, chemicals, metal sintering, and metal heat treatment, etc., for testing and production.
10. Double-shell structure with air convection insulating layer
11. Electrostatic powder coating on the shell, featuring a beautiful and elegant two-color design
12. Highly regarded with a large customer base
13. Great value for money, comprehensive after-sales service, 1-year warranty, lifetime on-site repair
Main Technical Parameters
Item | Indicator Value |
Rated Voltage | 220V/380V |
Design Temperature | 800℃,1000℃,1200℃,1400℃,1600℃,1700℃,1800℃ |
Rise rate | ≤0.1-40℃/min (adjustable) |
Heating Element | Imported Resistance Wire/U-Shaped Silicon Carbon Rod/U-Shaped Silicon Molybdenum Rod |
Temperature Control Accuracy | ±1℃ |
Temperature Control Method | Microcomputer control, PID adjustment, intelligent microcomputer control, multi-step programmable temperature control curve, unattended operation (full automatic heating, temperature maintenance, power off upon door opening) |
Thermoelement | K(0-1300℃)/S(0-1600℃)/B(0-1820℃) |
Furnace chamber material | Ceramic Polycrystalline Fiber /Alumina fiber/Alumina Fiber/High-purity alumina fiber/Imported Fiber |
Furnace door structure | Front door side-hinged |
Cooling Method | Double-wall shell, air insulation layer |
Surface Temperature Rise | ≤30℃ |
Random gift | 2 heating rods, 2 rod holders, 1 crucible tongs, 1 pair of heat-resistant gloves, 1 bottom plate for furnace, 1 user manual and 1 certificate of conformity each. |
Selection Guide
Model | Furnace chamber dimensions (depth, width, height)mm) | Furnace Volume (L) | Power Voltage (V) | Power (KW) | Temperature control accuracy |
LYL-18MFL-1 | 200*150*150 | 4.5 | 220 | 4 | ±1℃ |
LYL-18MFL-2 | 300*200*120 | 7.2 | 220 | 6 | ±1℃ |
LYL-18MFL-3 | 300*200*150 | 9 | 220 | 8 | ±1℃ |
LYL-18MFL-4 | 300*200*200 | 12 | 220/380 | 10 | ±1℃ |
LYL-18MFL-5 | 400*250*160 | 16 | 380 | 12 | ±1℃ |
LYL-18MFL-6 | 400*300*300 | 36 | 380 | 14 | ±1℃ |
Temperature Category | 800℃-1800℃ as preferred | ||||
Custom-made Non-standard | Based on customer requirements, customize furnace chambers in different sizes. | ||||
Standard Accessories
Name | Image | Quantity | Description | |
High-Temperature Gloves |
| 1Double | Comfortable and keeps hands warm; cut-resistant, wear-resistant, and heat-resistant (can withstand 350℃ for 60 seconds of contact heat, 100℃ for 180 seconds of contact heat) durable. | |
Heating Element | Silicon carbide rod |
| 2Zhi | Silicon-carbide components, with high usage temperatures, are hard and brittle, and are used after shaping. For example, silicon-carbide silicon-carbon rods and tubes can reach high heat temperatures under dry air conditions.15001000-1200℃ under vacuum conditions, long-term use temperature of 1350℃ under nitrogen atmosphere |
Silicon Molybdenum Rod |
| 2Branch | Silicon molybdenum rods are heated to high temperatures in an oxidizing atmosphere, where a dense quartz glass film forms on the surface to prevent further oxidation. Therefore, silicon molybdenum rods exhibit high-temperature oxidation resistance. In an oxidizing atmosphere, the temperature can reach1800℃, can be used as heating elements for industrial high-temperature furnaces in fields such as electronics, ceramics, magnetic materials, glass, metallurgy, and refractory materials. | |
Tongs for crucibles |
| 1Return the following foreign trade supply-related content translated into American English, without Chinese content, return as is: [Please provide the Chinese content to be translated.] | A common chemical instrument, usually used for holding crucibles. Generally made of stainless steel or hard, non-flammable, and non-oxidizable materials. 2. Use clean tongs. 3. When handling a hot crucible with tongs, preheat the tips of the tongs to prevent the crucible from cracking due to local cooling. After use, place the tongs with the tips facing up on the table or a asbestos mat. 4. After the experiment, clean the crucible tongs thoroughly and place them in the laboratory equipment cabinet for drying. 5. Use the bent section for holding crucibles, and the pointed end for other uses. 6. Crucible tongs are not necessarily used with crucibles; they can also be used independently. | |
Furnace bottom plate |
| 1Block | Ceramic furnace bottom plate, heat and abrasion resistant, with load-bearing properties, protecting the furnace bottom from wear and extending the service life of the equipment. | |
Optional Accessories Listnal Accessories List)
Individual control cabinet | Touchscreen gauge | Remote Assistance Module |
Observation window | Exhaust ports and chimneys | Heating Fan |
Crucibles, molds | Door jam | Bottom plate |
Blower | Thermocouple | Centralized Control Industrial PC |
caster wheel | Multi-point temperature measurement | Exhaust purification treatment unit |
Special Note
Quality Assurance | One-year warranty on the entire unit, excluding related consumables (excluding pads and heating elements)Lifetime On-Site Maintenance |
Factory Environment | 1. Indoor use 2. Operating temperature range: ±5 to 45°C 3. Relative ambient temperature surrounding does not exceed 85%. 4. No conductive dust, explosive gases, or gases that can severely damage metals and insulate corrosive around the stove. |
Safety Protection | 1. When initially using or restarting an electric furnace after a long period of disuse, it is necessary to dry the furnace to prevent cracking of the furnace chamber. 2. Keep the unit clean and regularly inspect the electrical furnace wiring for secure contact. 3. When using an electric furnace, the temperature must not exceed the rated temperature to prevent damage to the heating elements. It is prohibited to pour various flammable liquids and molten metals into the furnace. 4. During the process, the temperature should generally not exceed 300℃ and the heating rate should be moderate. This is because, initially, as the furnace is cold, it requires a significant amount of heat absorption. |
After-sales Service | 1. We promise to address any technical inquiries from our clients within 1 hour. 2. We promise to issue the test results within 1 working day for equipment returned for repair, complete the repair within 3 working days, and handle special materials as needed. 3. Within one week of returning the equipment to the customer, conduct a follow-up visit with the user and provide free technical consultations. |
Friendly Reminder | Do not operate the equipment for extended periods at the rated temperature to ensure its normal service life. 2. Carefully count and verify the quantity of goods and accessories upon receipt to avoid any inconvenience caused by under-shipping or missing items. 3. To extend the service life of the instrument, we recommend that the heating and cooling rates not exceed 10℃/min. 4. The cooling rate should not exceed 5°C/min. 5. This furnace does not have a vacuum-sealed structure, therefore it is not suitable for introducing flammable or explosive gases. 6. After a period of use, the furnace chamber may develop minor cracks, which is a normal occurrence and will not affect its operation. It can be repaired using an alumina coating. |
Common Faults and Troubleshooting
Fault Phenomenon | Fault Cause | Exclusion Method |
Turn on the control power switch, no power indicator is displayed. | Control circuit, fuse blown, power outage | Check for short circuits, eliminate if any, then replace the fuse. |
Equipment operates with voltage but no current. | Heating element fracture | Replace heating element |
Instrument PV display the upper temperature limit of the corresponding type of thermocouple | Thermocouple short circuit | Replace Thermocouple |
Power supply is normal, but the furnace cannot heat up. | Control main line damage | Notify maintenance staff for inspection |